DocumentCode
3202157
Title
Performance analysis of a Colorimeter designed with RGB color sensor
Author
Raja, Sivanantha A., Sr. ; Sankaranarayanan, Kavitha
Author_Institution
AC Coll. of Eng. & Tech., Karaikudi
fYear
2007
fDate
25-28 Nov. 2007
Firstpage
305
Lastpage
310
Abstract
The ldquoGold standardrdquo for testing blood glucose in Colorimeters is the measurement of glucose in a plasma sample obtained from a vein. This method involves a chemical reaction activated by an enzyme called Glucose Oxidase. Most of the Colorimeters used for the Glucose measurement in Clinical laboratories perform well in the 100-400 mg/dL range of blood glucose. Due to the requirement of highly monochromatic source and photo detector, an error of 5 to 20% is common in these Colorimeters. This paper aims to mitigate the error in measurement by modernizing the plasma glucose testing equipment i.e., Colorimeter with White Source and RGB Color Sensor through Color measurement technique for better clinical diagnosis. As presumed, experimental results show that this modified Colorimeter can perform well with better accuracy even above the Glucose concentration of 400 mg/dL.
Keywords
biomedical measurement; blood; chemical reactions; colorimeters; enzymes; RGB color sensor; blood glucose testing; chemical reaction; color measurement technique; colorimeter; enzyme; glucose oxidase; white source; Blood; Chemical sensors; Color; Performance analysis; Plasma chemistry; Plasma measurements; Plasma sources; Sugar; Testing; Veins; Absorbance; Chromaticity diagram; Color measurement; Colorimeter; Glucose measurement; RGB Color Sensor;
fLanguage
English
Publisher
ieee
Conference_Titel
Intelligent and Advanced Systems, 2007. ICIAS 2007. International Conference on
Conference_Location
Kuala Lumpur
Print_ISBN
978-1-4244-1355-3
Electronic_ISBN
978-1-4244-1356-0
Type
conf
DOI
10.1109/ICIAS.2007.4658396