• DocumentCode
    3214309
  • Title

    Field simulation of applicators for interstitial microwave hyperthermia

  • Author

    Schaller, G. ; Erb, J. ; Engelbrecht, Rainer ; Seegenschmiedt, H.M.

  • Author_Institution
    Lab. for High Frequency Technol., Erlangen-Nurnberg Univ., Germany
  • fYear
    1995
  • fDate
    16-20 May 1995
  • Firstpage
    307
  • Abstract
    The electromagnetic field near clinically applied coaxial microwave applicators for interstitial hyperthermia is calculated using a finite integration algorithm program. Simulations of E-field and SAR distributions of different conventional applicators in a muscle phantom are presented and compared with measurements. The applied calculation method allows also design optimization and improvement of more sophisticated applicator types, e.g. those using a triaxial technique.<>
  • Keywords
    biomedical equipment; hyperthermia; microwave heating; muscle; radiation therapy; E-field; SAR distributions; applicators; clinically applied coaxial microwave applicators; design optimization; electromagnetic field; field simulation; finite integration algorithm program; interstitial microwave hyperthermia; muscle phantom; triaxial technique; Antenna measurements; Applicators; Catheters; Coaxial components; Electromagnetic heating; Frequency; Hyperthermia; Imaging phantoms; Impedance; Muscles;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microwave Symposium Digest, 1995., IEEE MTT-S International
  • Conference_Location
    Orlando, FL, USA
  • ISSN
    0149-645X
  • Print_ISBN
    0-7803-2581-8
  • Type

    conf

  • DOI
    10.1109/MWSYM.1995.406045
  • Filename
    406045