The electromagnetic field near clinically applied coaxial microwave applicators for interstitial hyperthermia is calculated using a finite integration algorithm program. Simulations of E-field and SAR distributions of different conventional applicators in a muscle phantom are presented and compared with measurements. The applied calculation method allows also design optimization and improvement of more sophisticated applicator types, e.g. those using a triaxial technique.<>
Keywords
biomedical equipment; hyperthermia; microwave heating; muscle; radiation therapy; E-field; SAR distributions; applicators; clinically applied coaxial microwave applicators; design optimization; electromagnetic field; field simulation; finite integration algorithm program; interstitial microwave hyperthermia; muscle phantom; triaxial technique; Antenna measurements; Applicators; Catheters; Coaxial components; Electromagnetic heating; Frequency; Hyperthermia; Imaging phantoms; Impedance; Muscles;