DocumentCode :
323287
Title :
Characteristics of PbI2 crystal for use as a gamma-ray detector
Author :
Shoji, Tadayoshi ; Hitomi, Keitaro ; Tiba, Takamitu ; Suehiro, Teruo ; Hiratate, Yukio
Author_Institution :
Tohoku Inst. of Technol., Sendai, Japan
fYear :
1997
fDate :
9-15 Nov 1997
Firstpage :
804
Abstract :
We have been studying a γ-ray detector which uses a lead iodide PbI2 crystal. The detector fabricated from a PbI2 crystal grown by 10-pass zone travelling molten zone (TMZ) method. The μτproduct of this crystal are estimated to be about 1×10-6 cm2/V for electrons and about 6×10-7 cm2/V for holes. The resistivity of the PbI2 crystal is estimated to be about 1012 Ωcm. FWHM energy resolution of the detector for a photoelectric peak for the 59.5 keV γ-rays is found to be about 5 keV
Keywords :
gamma-ray detection; lead compounds; solid scintillation detectors; 59.5 keV; FWHM energy resolution; PbI2; crystal; gamma-ray detector; resistivity; Charge carrier processes; Charge carriers; Chemicals; Colloidal crystals; Electrodes; Etching; Gamma ray detectors; Photonic crystals; Pulse amplifiers; Voltage;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium, 1997. IEEE
Conference_Location :
Albuquerque, NM
ISSN :
1082-3654
Print_ISBN :
0-7803-4258-5
Type :
conf
DOI :
10.1109/NSSMIC.1997.672703
Filename :
672703
Link To Document :
بازگشت