    A statistical averaging method based on an extended Kalman filter is applied to digital sampling oscilloscope (DSO) measurements of modulated RF signals. Various measurements using a 48 Mbit/s 802.11a signal modulated onto a 1.9 GHz carrier are used to demonstrate significant performance enhancements as compared to traditional measurement techniques. The method presented shows that a general purpose high frequency DSO can replace more expensive and dedicated equipment for accurate characterization of microwave devices under modulated signal excitation
  • Keywords
    Kalman filters; UHF measurement; digital storage oscilloscopes; signal processing equipment; statistical analysis; 1.9 GHz; 48 Mbit/s; digital sampling oscilloscope; extended Kalman filter; microwave measurement; statistical averaging; statistical signal processing; Bandwidth; Digital modulation; Frequency measurement; Microwave measurements; Microwave technology; Microwave theory and techniques; Oscilloscopes; Performance evaluation; Radio frequency; Sampling methods; Digital sampling oscilloscope; microwave measurement techniques; statistical signal processing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Integrated Nonlinear Microwave and Millimeter-Wave Circuits, 2006 International Workshop on
  • Conference_Location
    Aveiro
  • Print_ISBN
    0-7803-9723-1
  • Electronic_ISBN
    0-7803-9723-1
  • Type

    conf

  • DOI
    10.1109/INMMIC.2006.283496
  • Filename
    4062247