• DocumentCode
    3296997
  • Title

    Insights into cardiac pacemaker and defibrillator revision/upgrades

  • Author

    Hanzlik, Josa ; Patel, Jasmine ; Kurtz, Steven ; Pavri, Behzad ; Greenspon, Arnold ; Ochoa, Jorge

  • Author_Institution
    Biomed. Eng., Drexel Univ., Philadelphia, PA, USA
  • fYear
    2011
  • fDate
    1-3 April 2011
  • Firstpage
    1
  • Lastpage
    2
  • Abstract
    Implantation of cardiac implantable electronic device (CIEDs), both pacemakers (PM) and implantable cardioverter defibrillators (ICDs) has dramatically increased in recent years. Cardiovascular disease causes an estimated 30% of all deaths worldwide. In a 14-month period, a consecutive series of 121 devices (76 ICDs and 45 PMs) were collected as part of an IRB-approved device retrieval program. The criteria for devices to be reprocessable are PMs with remaining projected longevity of greater than 4 years and ICDs with battery voltage greater than 2.95V were noted. Our data suggests that 17-20% of device removals in a teaching hospital may be suitable for re-use in countries with developing economies.
  • Keywords
    defibrillators; pacemakers; CIED; ICD; IRB-approved device retrieval program; cardiac implantable electronic device; cardiac pacemaker; cardiovascular disease; defibrillator revision; defibrillator upgrades; implantable cardioverter defibrillators; implantation; teaching hospital; Batteries; Correlation; Education; Hospitals; Implants; Lead; Pacemakers; Cardiac Implantable Electronic Device; Cardioverter-defibrillator; Pacemaker; Retrieval Analysis;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Bioengineering Conference (NEBEC), 2011 IEEE 37th Annual Northeast
  • Conference_Location
    Troy, NY
  • ISSN
    2160-7001
  • Print_ISBN
    978-1-61284-827-3
  • Type

    conf

  • DOI
    10.1109/NEBC.2011.5778519
  • Filename
    5778519