  • Abstract
    A new inductive coupling based wireless system is being developed to monitor the condition status of aortic stent-grafts. It relies on the measure of the stent-graft´s outer pressure using a capacitive sensor placed in a LC resonant circuit. The work presented herein addresses the testing of the LC sensor circuit to diagnose whether observed pressure deviations are due to defects occurring in the sensor´s inductor and capacitor or to an actual degradation of the stent-graft.
  • Keywords
    LC circuits; biomedical electronics; biomedical telemetry; capacitive sensors; capacitors; cardiology; fault diagnosis; inductors; pressure measurement; pressure sensors; stents; LC resonant circuit; aortic stent-grafts; capacitive sensor; capacitor; fault detection system; inductive coupling; inductor; outer pressure measurement; stent-graft endoleakage monitor; telemetry system; Capacitors; Frequency measurement; Impedance; Inductors; Pressure measurement; Resonant frequency; Sensors; fault detection; monitoring; stent-graft;