New attenuation correction for the HRRT using transmission scatter correction and total variation regularization
Author :
Sibomana, Merence ; Keller, Sune Hogild ; Svarer, Claus ; Olesen, Oline Vinter ; Andersen, Flemming ; Holm, Soren ; Hjgaard, L.
fYear :
2009
fDate :
Oct. 24 2009-Nov. 1 2009
Firstpage :
3284
Lastpage :
3286
Abstract :
In the standard software for the Siemens HRRT PET scanner the most commonly used segmentation in the ¿-map reconstruction for human brain scans is MAP-TR. Problems with bias in the lower cerebellum and pons in HRRT brain images have been reported. The main source of the problem is poor bone / soft tissue segmentation in these regions and the lack of scatter correction in the ¿-map reconstruction. In this paper we describe and validate the new TXTV segmentation method (included in the HRRTU 1.0 and 1.1 user software) aimed at solving the bias problem.
Keywords :
brain; positron emission tomography; HRRT brain images; Siemens HRRT PET scanner; human brain scans; lower cerebellum; total variation regularization; transmission scatter correction; Attenuation; Biological tissues; Bones; Brain; Humans; Image reconstruction; Image segmentation; Positron emission tomography; Scattering; Software standards;