• DocumentCode
    333338
  • Title

    Frameless stereotactic localization and multimodal image registration using DSA/CT/MRI

  • Author

    Fei, Baowi ; Zhuang, Tiange ; Hu, Jie ; Zhou, Fanmin

  • Author_Institution
    Dept. of Biomed. Eng., Shanghai Jiaotong Univ., China
  • Volume
    2
  • fYear
    1998
  • fDate
    29 Oct-1 Nov 1998
  • Firstpage
    683
  • Abstract
    A novel frameless stereotactic localization method and multimodal image registration techniques for computer assisted surgery are presented. Using four external markers, the position of brain tissue can be calculated by two different DSA projection images, CT slices and MRI slices respectively. The different anatomical information (tone, soft tissue and vessel) obtained from CT, MRI and DSA can be registered together. Using DSA, the locating accuracy of phantom experiment is 0.5 mm, the accuracy of skull experiment is 2.0 mm. The registration accuracy of DSA/CT images of skull is 2.0-2.5 mm. The volumetric representation of registered DSA and MRI images of human head shows satisfying results
  • Keywords
    biomedical MRI; brain; computerised tomography; diagnostic radiography; image registration; medical image processing; surgery; CT; DSA; MRI; brain tissue position; computer assisted surgery; digital subtraction angiography; external markers; frameless stereotactic localization; human head; magnetic resonance imaging; medical diagnostic imaging; multimodal image registration; skull experiment accuracy; Biological tissues; Brain; Computed tomography; Head; Humans; Image registration; Imaging phantoms; Magnetic resonance imaging; Skull; Surgery;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1998. Proceedings of the 20th Annual International Conference of the IEEE
  • Conference_Location
    Hong Kong
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-5164-9
  • Type

    conf

  • DOI
    10.1109/IEMBS.1998.745510
  • Filename
    745510