• DocumentCode
    333519
  • Title

    A non invasive rechargeable cardiac pacemaker battery system with a transcutaneous energy transformer

  • Author

    Nishimura, Toshi Hiro ; Eguchi, Tetsuji ; Kubota, Akira ; Hatori, Mitsutoshi ; Saito, Masao

  • Author_Institution
    Dept. of Electr. & Electron. Eng., Oita Univ., Japan
  • fYear
    1998
  • fDate
    29 Oct-1 Nov 1998
  • Firstpage
    432
  • Abstract
    The improvement method of a flat secondary transcutaneous transformer for a resonant converter that transfers several W of power across a large variable human skin gap is presented and built. The secondary transformer is piled, and the coupling coefficient, conversion efficiency and temperature rise are measured with an experimental set up. The coupling coefficient and the conversion efficiency are increased compared with the conventional transformer. The temperature rise is decreased compared with the conventional transformer
  • Keywords
    DC-DC power convertors; battery chargers; pacemakers; prosthetic power supplies; resonant power convertors; conversion efficiency; coupling coefficient; flat secondary transcutaneous transformer; large variable human skin gap; noninvasive system; rechargeable cardiac pacemaker battery system; resonant converter; temperature rise; Amorphous materials; Batteries; Couplings; DC-DC power converters; Frequency conversion; Pacemakers; Power supplies; Resonance; Skin; Transformer cores;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1998. Proceedings of the 20th Annual International Conference of the IEEE
  • Conference_Location
    Hong Kong
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-5164-9
  • Type

    conf

  • DOI
    10.1109/IEMBS.1998.745938
  • Filename
    745938