DocumentCode
333526
Title
Automatic control of assist flow in totally implantable left ventricular assist device using variable pump rate
Author
Choi, Jonghoon ; Choi, Jaesoon ; Jo, Youngho ; Kim, Wukeun ; Min, Byunggoo
Author_Institution
Interdisciplinary Program of Med. & Biol. Eng., Seoul Nat. Univ., South Korea
fYear
1998
fDate
29 Oct-1 Nov 1998
Firstpage
462
Abstract
In this study, the automatic control algorithm in implantable left ventricular assist device to regulate cardiac output to the extent of preventing atrial-collapse and the high left atrial pressure (LAP) is proposed. Intersac pressure (ISP) waveform which reflects the status of left atrial pressure is used in automatic control. By analyzing the relationship between LAP and the negative peak of ISP, the equation for making the reference of PI-controller was set. By using the reference equation the proposed automatic controller which regulates the assist flow was designed. Through mock circulatory experiments and animal experiment, the automatic control algorithm was evaluated
Keywords
biocontrol; cardiovascular system; electrohydraulic control equipment; flow control; prosthetics; pumps; two-term control; PI-controller reference; animal experiment; assist flow; atrial-collapse prevention; automatic control; cardiac output regulation; high left atrial pressure; intersac pressure waveform; left ventricular assist device; mock circulatory experiments; physiological control; totally implantable; variable pump rate; Algorithm design and analysis; Animals; Automatic control; Biomedical engineering; Blood; Educational institutions; Electronic mail; Engineering in medicine and biology; Equations; Pi control;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1998. Proceedings of the 20th Annual International Conference of the IEEE
Conference_Location
Hong Kong
ISSN
1094-687X
Print_ISBN
0-7803-5164-9
Type
conf
DOI
10.1109/IEMBS.1998.745947
Filename
745947