Abstract
A first prototype of a high resolution small animal PET scanner based on continuous LYSO scintillator crystals and silicon photomultiplier matrices has been developed at IFIC-Valencia, in collaboration with the University of Pisa and INFN Pisa. The prototype consists of two detector heads attached to a rotating stage. Each head is composed of a continuous 12 mm × 12 mm × 5 mm LYSO crystal painted white, coupled to a monolithic SiPM matrix. The matrices, developed at FBK-irst, are composed of 8×8 SiPM elements of 1.5 mm×1.4 mm size in a common substrate. The full characterization of the detector heads with different types of crystals has been carried out, and a method for determining the interaction position of the gamma-rays in the crystals has been successfully employed. Tomographic data have been acquired with the prototype, and images of different source distributions have been reconstructed with the Maximum Likelihood Expectation Maximization algorithm (MLEM). A FWHM close to 1 mm is obtained for one and two point-like sources.
Keywords
data acquisition; expectation-maximisation algorithm; gamma-ray detection; image reconstruction; image resolution; image scanners; medical image processing; photomultipliers; positron emission tomography; silicon radiation detectors; solid scintillation detectors; IFIC-Valencia; INFN Pisa; University of Pisa; continuous LYSO scintillator crystals; data acquisition; full detector characterization; gamma-ray interacting position determination; high resolution small animal PET scanner prototype; maximum likelihood expectation maximization algorithm; monolithic SiPM matrices; point-like sources; silicon photomultiplier matrices; tomographic data; Europe; Image resolution; Silicon photomultipliers; continuous crystals; multicell G-APD; small animal PET;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
Conference_Location
Valencia
ISSN
1082-3654
Print_ISBN
978-1-4673-0118-3
Type
conf
DOI
10.1109/NSSMIC.2011.6153684
Filename
6153684
Link To Document