• DocumentCode
    3344022
  • Title

    A method for measuring the sub-pixel light distribution of scintillation detectors with digital SiPMs

  • Author

    Düppenbecker, Peter Michael ; Haagen, Rainald ; Lodomez, Sarah ; Marsden, Paul K. ; Schulz, Volkmar

  • Author_Institution
    Philips Res. Eur., Aachen, Germany
  • fYear
    2011
  • fDate
    23-29 Oct. 2011
  • Firstpage
    2301
  • Lastpage
    2302
  • Abstract
    One of the main features of the digital silicon photomultiplier (dSiPM) is the ability to enable and disable individual cells of the sensor by reprogramming. This feature is normally used to disable noisy cells with high dark count rates to reduce the noise and dead time of the detector. In this work we show how this feature can be used to measure the light distribution of scintillation detectors with sub-pixel resolution by systematically enabling and disabling certain areas of the sensor.
  • Keywords
    biomedical equipment; photomultipliers; positron emission tomography; scintillation counters; silicon radiation detectors; PET detector; dark count rate; digital silicon photomultiplier; noisy cells; scintillation detectors; subpixel light distribution;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
  • Conference_Location
    Valencia
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4673-0118-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2011.6153866
  • Filename
    6153866