  • Abstract
    Light scattering technique is used to measuring the blood cell size in many types of Hematology Analyzers, but How to choose the detection angle of scattered light affect the efficiency of the instrument count. In this paper, axial lighting model is established. Mathematical description of luminous flux and particle size of cell is given. And through numerical calculation, the main factors counting efficiency are analysis.
  • Keywords
    biological techniques; blood; cellular biophysics; light scattering; particle size; Hematology Analyzers; axial lighting model; blood cell size; detector arrangement; forward light scattering; particle size; single cell counter; Cells (biology); Counting circuits; Instruments; Light scattering; Mie scattering; Optical scattering; Particle scattering; Red blood cells; Size measurement; Volume measurement; Light scattering; Numerical Calculation; luminous flux;