DocumentCode
3353956
Title
Sequential discrimination of atrial and ventricular tachyarrhythmias
Author
Mears, D. ; Pan, K. ; Xin-rong, G. ; Thakor, N.
Author_Institution
Dept. of Biomed. Eng., Johns Hopkins Med. Sch., Baltimore, MD, USA
fYear
1990
fDate
23-26 Sep 1990
Firstpage
107
Lastpage
110
Abstract
Described is a sequential hypothesis testing algorithm capable of discriminating supraventricular arrhythmias from ventricular arrhythmias. The algorithm uses simultaneous recordings of pure atrial and pure ventricular electrograms obtained using dual chamber endocardial electrodes on a single catheter. Adaptive impulse filtering is used to subtract the ventricular component from the atrial lead signal to get a pure atrial electrogram. Probability distributions of the atrial and ventricular inter-beat interval for various arrhythmias are calculated. A sequential hypothesis testing (SHT) algorithm is used to determine which distribution the rhythm belongs to. The advantage of this method is that it allows preselection of the false positive and false negative error probabilities. The tradeoff for selecting lower error rates is a longer time required to reach a decision. The SHT algorithm is used to discriminate between normal and SVT rhythms in 20 data segments. The 9 normal rhythms and 10 SVT rhythms were classified correctly within 3 seconds
Keywords
electrocardiography; waveform analysis; 3 s; adaptive impulse filtering; atrial tachyarrhythmias; dual chamber endocardial electrodes; error rates; inter-beat interval; probability distribution; sequential hypothesis testing algorithm; tachyarrhythmias discrimination; ventricular tachyarrhythmias; Adaptive filters; Catheters; Electrocardiography; Electrodes; Error analysis; Filtering; Frequency; Probability distribution; Rhythm; Sequential analysis;
fLanguage
English
Publisher
ieee
Conference_Titel
Computers in Cardiology 1990, Proceedings.
Conference_Location
Chicago, IL
Print_ISBN
0-8186-2225-3
Type
conf
DOI
10.1109/CIC.1990.144175
Filename
144175
Link To Document