DocumentCode
335991
Title
Validation of RF catheter ablation models using basket electrodes
Author
Cao, Hong ; Tungjitkusolmun, Supan ; Tsai, Jang-Zem ; Webster, John G.
Author_Institution
Dept. of Electr. & Comput. Eng., Wisconsin Univ., Madison, WI, USA
Volume
1
fYear
1997
fDate
30 Oct-2 Nov 1997
Firstpage
395
Abstract
The authors are using basket electrodes to validate electrical-thermal computer models of RF catheter ablation for the cure of tachycardias. They have measured electrical characteristics of the basket electrode and shown that accuracy degrades above 100 kHz. They present some preliminary measurements and suggest considerations for the validation of the model for RF catheter ablation
Keywords
bioelectric phenomena; biomedical electrodes; biothermics; cardiology; patient treatment; physiological models; 100 kHz; RF catheter ablation models validation; basket electrodes; cardiac rhythm disorders; electrical characteristics; electrical-thermal computer models; tachycardias cure; therapeutic technique; Blood; Catheters; Electric variables measurement; Electrodes; Electromagnetic heating; Heart; Laplace equations; Radio frequency; Solid modeling; Temperature;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1997. Proceedings of the 19th Annual International Conference of the IEEE
Conference_Location
Chicago, IL
ISSN
1094-687X
Print_ISBN
0-7803-4262-3
Type
conf
DOI
10.1109/IEMBS.1997.754561
Filename