DocumentCode :
3365077
Title :
Volumetric imaging using 2D capacitive micromachined ultrasonic transducer arrays (CMUTs): initial results
Author :
Oralkan, Ömer ; Ergun, A. Sanli ; Cheng, Ching-Hsiang ; Johnson, Jeremy A. ; Karaman, Mustafa ; Lee, Thomas H. ; Khuri-Yakub, Butrus T.
Author_Institution :
Edward L. Ginzton Lab., Stanford Univ., USA
Volume :
2
fYear :
2002
fDate :
8-11 Oct. 2002
Firstpage :
1083
Abstract :
This paper presents the first volumetric images obtained using a 2D CMUT array with through-wafer via interconnects. An 8×6-element portion of a 32×64-element array flip-chip bonded onto a glass fanout chip was used in the experiments. This study experimentally demonstrates that 2D CMUT arrays can be fabricated with high yield using silicon micromachining processes, individual electrical connections can be provided using through-wafer interconnects, and the flip-chip bonding technique can be used to integrate the dense 2D arrays with electronic circuits for practical imaging applications.
Keywords :
flip-chip devices; interconnections; micromachining; silicon; ultrasonic imaging; ultrasonic transducer arrays; 2D CMUT array; 2D capacitive US transducer arrays; Si; Si micromachining processes; dense 2D arrays; flip-chip bonding; glass fanout chip; micromachined ultrasonic transducer arrays; through-wafer via interconnects; volumetric imaging applications; Bonding; Fabrication; Integrated circuit interconnections; Laboratories; Phased arrays; Silicon; Spatial resolution; Ultrasonic imaging; Ultrasonic transducer arrays; Ultrasonic transducers;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 2002. Proceedings. 2002 IEEE
ISSN :
1051-0117
Print_ISBN :
0-7803-7582-3
Type :
conf
DOI :
10.1109/ULTSYM.2002.1192483
Filename :
1192483
Link To Document :