  • Abstract
    A nerve cuff electrode interface capable of both stimulating and recording from a nerve is described. The interface also rejects the EMG contamination in the recordings using reactive components without adding noise to the ENG signal. A transformer is added to the design for noise matching and the signal-to-noise ratio improvement is evaluated for a specific amplifier (AMP-O1)
  • Keywords
    biomedical electrodes; biomedical electronics; electromyography; neurophysiology; EMG contamination rejection; amplifier; cuff electrodes; nerve recording; nerve stimulation; noise matching; signal-to-noise ratio improvement; transformer; Biomedical electrodes; Contamination; Electromyography; Impedance; Noise cancellation; Noise figure; Noise generators; Signal design; Signal to noise ratio; Student members;