DocumentCode
3374387
Title
Measurement of left ventricular ejection fraction (EF) by densitometry from digital subtraction angiography
Author
Machnig, Th ; Eicker, B. ; Barth, Kurt ; Lehmkuhl, H. ; Bachmann, K.
Author_Institution
Div. of Cardiology, Erlangen Univ., Germany
fYear
1990
fDate
23-26 Sep 1990
Firstpage
589
Lastpage
592
Abstract
The densitometric and geometric method for estimating the left ventricular ejection fraction (EF) are compared and the accuracy and reproducibility of densitometry in patients and casts are assessed. EF calculations for the densitometric method are in a better agreement with the simulated EF in the heart casts than the area-length method. The intra- and inter-observer variability is remarkably lower for the densitometry than for the area-length method. The data indicate that the densitometric calculation of EF is a more reliable method than the area-length method, because densitometry is independent of geometric assumptions
Keywords
biomedical measurement; cardiology; densitometry; diagnostic radiography; area-length method; densitometry; digital subtraction angiography; geometric method; heart casts; left ventricular ejection fraction measurement; patients; Angiography; Biomedical engineering; Biomedical imaging; Cardiology; Computational modeling; Density measurement; Heart; Imaging phantoms; Shape; X-ray imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Computers in Cardiology 1990, Proceedings.
Conference_Location
Chicago, IL
Print_ISBN
0-8186-2225-3
Type
conf
DOI
10.1109/CIC.1990.144287
Filename
144287