• DocumentCode
    3380098
  • Title

    New Column Designs for MicroGC

  • Author

    Radadia, Adarsh D. ; Masel, Richard I. ; Shannon, Mark A.

  • Author_Institution
    Univ. of Illinois at Urbana-Champaign, Urbana
  • fYear
    2007
  • fDate
    10-14 June 2007
  • Firstpage
    2011
  • Lastpage
    2014
  • Abstract
    In this paper, we present some new column designs for micro-gas chromatography (muGC) columns. Serpentine columns are compared to commonly used spiral columns and are found to give relatively lower height equivalent to theoretical plate (HETP) values. Turns are a major dispersion source for serpentine configuration. We found that sinusoidal compensation structures following the turns give lowest dispersion amongst the several tested turn geometries for serpentine configuration.
  • Keywords
    chromatography; gas sensors; microsensors; column design; micro-gas chromatography column; microGC; serpentine column; serpentine configuration; sinusoidal compensation structure; Actuators; Design engineering; Electrokinetics; Fabrication; Geometry; Microchannel; Silicon; Solid state circuits; Spirals; Testing; dispersion in microchannels; microGC; microcolumns;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Solid-State Sensors, Actuators and Microsystems Conference, 2007. TRANSDUCERS 2007. International
  • Conference_Location
    Lyon
  • Print_ISBN
    1-4244-0842-3
  • Electronic_ISBN
    1-4244-0842-3
  • Type

    conf

  • DOI
    10.1109/SENSOR.2007.4300557
  • Filename
    4300557