Title
New Column Designs for MicroGC
Author
Radadia, Adarsh D. ; Masel, Richard I. ; Shannon, Mark A.
Author_Institution
Univ. of Illinois at Urbana-Champaign, Urbana
fYear
2007
fDate
10-14 June 2007
Firstpage
2011
Lastpage
2014
Abstract
In this paper, we present some new column designs for micro-gas chromatography (muGC) columns. Serpentine columns are compared to commonly used spiral columns and are found to give relatively lower height equivalent to theoretical plate (HETP) values. Turns are a major dispersion source for serpentine configuration. We found that sinusoidal compensation structures following the turns give lowest dispersion amongst the several tested turn geometries for serpentine configuration.
Keywords
chromatography; gas sensors; microsensors; column design; micro-gas chromatography column; microGC; serpentine column; serpentine configuration; sinusoidal compensation structure; Actuators; Design engineering; Electrokinetics; Fabrication; Geometry; Microchannel; Silicon; Solid state circuits; Spirals; Testing; dispersion in microchannels; microGC; microcolumns;