    A high-performance adaptive ENG amplifier

  • Author

    Triantis, Iasonas F. ; Demosthenous, Andreas

  • Author_Institution
    Dept. of Electron. & Electr. Eng., Univ. Coll. London, UK
  • fYear
    2004
  • fDate
    1-3 Dec. 2004
  • Abstract
    This paper describes an integrated adaptive amplifier for receding electroneurogram (ENG) from tripolar cuff electrodes. The amplifier can automatically compensate for cuff imbalance, thus offering a fully implantable solution to this problem. The circuit was fabricated in 0.8 μm BiCMOS technology with an active area of 0.7 mm . Measured results showed that the ENG amplifier has a cuff imbalance correction range of > ±40%, and a signal-to-interference ratio of about 2/1 (6 dB) even for ±40% imbalance.
  • Keywords
    BiCMOS integrated circuits; bioelectric potentials; microelectrodes; neurophysiology; operational amplifiers; preamplifiers; prosthetics; BiCMOS technology; cuff imbalance; electroneurogram; high-performance adaptive ENG amplifier; tripolar cuff electrodes; BiCMOS integrated circuits; Circuit synthesis; Differential amplifiers; Educational institutions; Electric variables control; Electrodes; Electromyography; Interference; Preamplifiers; Voltage;