DocumentCode :
3441514
Title :
Practical ultrasonic spectrometric measurement of solution concentrations by a tracking technique
Author :
Dion, J.L. ; Barwicz, A. ; Bouchard, J.
Author_Institution :
Groupe de Recherche en Electron. Ind., Quebec Univ., Trois-Rivieres, Que., Canada
fYear :
1988
fDate :
2-5 Oct 1988
Firstpage :
529
Abstract :
The authors designed and tested a practical measurement technique for solutions, based on a resonator with enclosed piezoceramics which can operate at a resonant frequency mode around 500 kHz. The resonance can be tracked automatically by means of a phase-locked loop. The authors report measurements made on water-methanol mixtures, showing that concentration can be expressed as a function of temperature and resonant frequency. Concentration could be evaluated with a precision of about 0.04% which depends essentially on the temperature measurement. Various milk mixtures have also been measured, to check the evaluation of fat, solid nonfat, and protein concentrations from measurements at two different temperatures. The system was designed to be eventually portable for field measurements
Keywords :
solutions; ultrasonic measurement; 500 kHz; milk mixtures; phase-locked loop; piezoceramics; resonant frequency mode; solution concentrations; temperature measurement; tracking technique; ultrasonic spectrometric measurement; Measurement techniques; Phase locked loops; Piezoelectric materials; Resonance; Resonant frequency; Spectroscopy; Temperature; Testing; Tracking loops; Ultrasonic variables measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 1988. Proceedings., IEEE 1988
Conference_Location :
Chicago, IL
Type :
conf
DOI :
10.1109/ULTSYM.1988.49433
Filename :
49433
Link To Document :
بازگشت