• DocumentCode
    344367
  • Title

    Magnetic field calculation for a 13 MeV PET cyclotron

  • Author

    Yang, W.Y. ; Yoon, M. ; Oh, S. ; Chai, J.S. ; Ha, J.H.

  • Author_Institution
    Dept. of Phys., POSTECH, South Korea
  • Volume
    4
  • fYear
    1999
  • fDate
    1999
  • Firstpage
    2265
  • Abstract
    A design study of the main magnet for a 13 MeV cyclotron has been carried out in a joint collaboration between the Korea Cancer Center Hospital (KCCH) and POSTECH. A maximum energy of 13 MeV has been chosen to produce radioisotopes such as 18F. There are four magnet sectors, each with radial-ridged shape. Maximum magnetic fields are 1.85 T and 0.48 T at hill and valley centers, respectively. The total size of the cyclotron is less than 2 m in diameter. The program TOSCA has been utilized for the field calculation and optimization. In this paper, we describe design parameters of the 13 MeV PET cyclotron, with emphasis on the magnetic field shape and the beam optics calculation
  • Keywords
    accelerator magnets; cyclotrons; ion accelerators; ion optics; magnetic fields; positron emission tomography; 0.48 T; 1.85 T; 13 MeV; 18F; PET cyclotron; TOSCA; beam optics; magnet; magnetic field shape; Cancer; Collaboration; Cyclotrons; Hospitals; Magnetic fields; Optical beams; Optical design; Positron emission tomography; Radioactive materials; Shape;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Particle Accelerator Conference, 1999. Proceedings of the 1999
  • Conference_Location
    New York, NY
  • Print_ISBN
    0-7803-5573-3
  • Type

    conf

  • DOI
    10.1109/PAC.1999.792654
  • Filename
    792654