Title :
Phantoms for elastography
Author :
Hall, Timothy J. ; Bilgen, Mehmet ; Insana, Michael F. ; Chaturvedi, Pooja
Author_Institution :
Dept. of Radiol., Univ. of Kansas Med. Center, Kansas City, KS, USA
Abstract :
Several viscoelastic properties of gelatin materials used in sonography and elastography phantoms are reported. Also reported is a device that was designed and built to make these measurements. Measurement precision with this device is typically within 1%. Finite element analysis was used to study boundary conditions and compressor size (compared to sample size) effects on the estimated elastic modulus. Results of these measurements indicate that gelatin gels are a good choice for elastography phantoms, but that the stiffness of these materials increases with time
Keywords :
biomechanics; biomedical equipment; biomedical ultrasonics; elasticity; finite element analysis; gelatin; viscoelasticity; boundary conditions; compressor size effects; elastography phantoms; material stiffness; measurement precision; medical diagnostic imaging; palpation; sample size effects; sonography; stiffness; tissue elasticity; Acoustic pulses; Acoustic scattering; Biological materials; Biological tissues; Elasticity; Imaging phantoms; Manufacturing; Stress; Ultrasonography; Viscosity;
Conference_Titel :
Ultrasonics Symposium, 1996. Proceedings., 1996 IEEE
Conference_Location :
San Antonio, TX
Print_ISBN :
0-7803-3615-1
DOI :
10.1109/ULTSYM.1996.584204