Title :
An improved multicrystal 2-D BGO detector for PET
Author :
Rogers, J.G. ; Taylor, A.J. ; Rahimi, M.F. ; Nutt, R. ; Andreaco, M. ; Williams, C.W.
Author_Institution :
TRIUMF, Vancouver, BC, Canada
Abstract :
The authors evaluate and compare two 2-D array detectors for PET (positron emission tomography). Both consist of an 8*8 array of small BGO crystals coupled to a 2*2 array of photomultiplier tubes. The depth of the crystals is 3 cm in one detector and 2 cm in the other. The 2-cm detector is obviously superior in terms of material costs, but it is also superior in energy resolution per crystal, and in its ability to identify clearly the crystal containing the primary interaction. The authors present a flexible and robust algorithm for crystal identification in such array detectors. The prospect of obtaining still better spatial resolution from such block detectors, with increased numbers of crystals, is discussed.<>
Keywords :
biomedical equipment; computerised tomography; radioisotope scanning and imaging; 2 cm; 3 cm; block detectors; crystal identification; energy resolution; flexible robust algorithm; material costs; medical diagnostic imaging; medical instrumentation; multicrystal 2D BGO detector; nuclear medicine; photomultiplier tubes; positron emission tomography; Costs; Crystalline materials; Crystals; Detectors; Energy resolution; Photomultipliers; Positron emission tomography; Robustness; Sensor arrays; Spatial resolution;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference, 1991., Conference Record of the 1991 IEEE
Conference_Location :
Santa Fe, NM, USA
Print_ISBN :
0-7803-0513-2
DOI :
10.1109/NSSMIC.1991.259186