DocumentCode :
3461658
Title :
Design and measurement of a WR-28 calorimeter
Author :
Cui, Xiaohai ; Liu, Xinmeng ; Li, Yong ; Xu, Hui ; Xing, Haixia
Author_Institution :
Nat. Inst. of Metrol. of China, China
fYear :
2010
fDate :
13-18 June 2010
Firstpage :
773
Lastpage :
774
Abstract :
We have designed and manufactured a new WR-28 (26.5-40GHz) rectangular waveguide calorimeter. The former WR-28 calorimeter was setup during 1980s, many parts are not work, and especially the mount is not available. So the new calorimeter needed. The calorimeter is designed as twin line structure and put in two layers copper barrel. The outer one is temperature controlled to reduce the effect of the change of ambient temperature. The temperature wave in inner barrel is less than 50 micro degrees within 24 hours. General used microwave load be used as the terminator, other than special designed mount. Resistor heater is tightly tied to load which can provide DC bias power. A commercial bolometric sensor embedded into a hole in the waveguide flange with a Type IV power meter is used to detect the DC substitution power.
Keywords :
bolometers; calorimeters; power meters; rectangular waveguides; temperature control; DC bias power; RF terminator; WR-28 rectangular waveguide calorimeter; ambient temperature; bolometric sensor; frequency 26.5 GHz to 40 GHz; microwave load; resistor heater; temperature control; twin line structure; two layer copper barrel; type IV power meter; Bridge circuits; Copper; Electromagnetic measurements; Electromagnetic waveguides; Flanges; Manufacturing; Radio frequency; Temperature control; Temperature sensors; Thermistors;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements (CPEM), 2010 Conference on
Conference_Location :
Daejeon
Print_ISBN :
978-1-4244-6795-2
Type :
conf
DOI :
10.1109/CPEM.2010.5543394
Filename :
5543394
Link To Document :
بازگشت