DocumentCode :
3461774
Title :
The status of the ICR/RAL, BaF2-TMAE positron camera
Author :
Wells, K. ; Ott, R.J. ; Suckling, J. ; Bateman, J.E. ; Stephenson, R. ; Connolly, J.F.
Author_Institution :
Inst. of Cancer Res., R. Marsden Hospital, Belmont, UK
fYear :
1991
fDate :
2-9 Nov. 1991
Firstpage :
1668
Abstract :
A novel BaF/sub 2/-TMAE (tetrakis dimethylaminoethylene) detector has been developed which can simultaneously achieve 5-ns timing, 5-mm spatial resolution, and 20% detection efficiency. The chamber can operate in a stable condition for periods of several hours. Results of projection imaging are also presented. The projection images obtained with the 100-cm/sup 2/ sensitive area detector represent the first occasion a distributed source has been used in conjunction with a BaF/sub 2/-TMAE detector for positron imaging.<>
Keywords :
biomedical equipment; cameras; computerised tomography; radioisotope scanning and imaging; ICR/RAL, BaF/sub 2/-TMAE positron camera; detection efficiency; detector; distributed source; medical diagnostic imaging; medical instrumentation; nuclear medicine; positon emission tomography; projection imaging; spatial resolution; tetrakis dimethylaminoethylene; Cameras; Circuits; Crystals; Detectors; Electrons; Insulation; Positrons; Spatial resolution; Vacuum systems; Wire;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference, 1991., Conference Record of the 1991 IEEE
Conference_Location :
Santa Fe, NM, USA
Print_ISBN :
0-7803-0513-2
Type :
conf
DOI :
10.1109/NSSMIC.1991.259198
Filename :
259198
Link To Document :