DocumentCode :
3462744
Title :
Non-stationary filtering technique for improved cone beam SPECT
Author :
Cao, Z.J. ; Tsui, B.M.W.
Author_Institution :
North Carolina Univ., Chapel Hill, NC, USA
fYear :
1991
fDate :
2-9 Nov. 1991
Firstpage :
1938
Abstract :
A truly 3-D reconstruction formulation is derived for the single-orbit cone beam SPECT (single photon emission computed tomography) based on the 3-D Fourier slice theorem. In the formulation, a nonstationary filter is derived which depends on the distance from the central plane of the cone beam. This filter is applied to the 2-D projection data in directions along and normal to the axis of rotation. The nonstationary filtering technique was evaluated using both computer simulation and phantom measurements. Significant improvement in image quality was demonstrated in terms of decreased artifacts along the axis of rotation at the cost of a fivefold increase in processing time. The quality of the reconstructed image from the 3-D algorithm is superior to that from the Feldkamp algorithm, especially near the ends of the image along the axis of rotation.<>
Keywords :
computerised tomography; radioisotope scanning and imaging; 3D Fourier slice theorem; 3D reconstruction formulation; Feldkamp algorithm; artifacts; computer simulation; cone beam SPECT; image quality; medical diagnostic imaging; nuclear medicine; phantom measurements; processing time; Biomedical engineering; Collimators; Filtering; Filters; Geometry; Image quality; Image reconstruction; Radiology; Reconstruction algorithms; Tomography;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference, 1991., Conference Record of the 1991 IEEE
Conference_Location :
Santa Fe, NM, USA
Print_ISBN :
0-7803-0513-2
Type :
conf
DOI :
10.1109/NSSMIC.1991.259255
Filename :
259255