Loading effect of the AC-DC transfer standard for low level AC voltage standard
Author
Kwon, Sung-Won ; Mun-Seog Kim ; Kim, Mun-Seog ; Jung, J.K.
Author_Institution
Korea Res. Inst. of Stand. & Sci., Daejeon, South Korea
fYear
2010
fDate
13-18 June 2010
Firstpage
173
Lastpage
174
Abstract
AC-DC transfer standard and micro-potentiometer are widely used for AC voltage standard at low level. Input impedance and 50-Ω load dependency of the standard are discussed for the ac voltage application below 200 mV and 1 MHz. The load dependency of the unit reaches 385 μV/V at 1 MHz, and this loading error should be corrected when calibrate the 50-Ω AC source using the standard.
Keywords
calibration; electric impedance measurement; measurement standards; potentiometers; AC-DC transfer standard; calibration; frequency 1 MHz; input impedance; load dependency; low level AC voltage standard; micropotentiometer; resistance 50 ohm; Calibration; Capacitance; Electrical resistance measurement; Electromagnetic measurements; Frequency measurement; Impedance measurement; Low voltage; Measurement standards; Resistors; Thermal resistance;