• DocumentCode
    347102
  • Title

    Cuff configuration for reduction of interfering signals in nerve cuff recordings

  • Author

    Andreasen, Lotte N S ; Struijk, Johannes J.

  • Author_Institution
    Center for Sensory Motor Interaction, Aalborg Univ., Denmark
  • Volume
    1
  • fYear
    1999
  • fDate
    1999
  • Abstract
    A new cuff configuration for reduction of interfering signals in nerve cuff recordings is introduced. It is based on the traditional tripolar configuration, but has additional end-electrodes which are short circuited and floating. The additional end-electrodes reduced the interfering signal by a factor of 4 to 1, and when platinized by a factor of 13 to 1
  • Keywords
    bioelectric potentials; biomedical electrodes; interference (signal); neurophysiology; Pt; cuff configuration; end-electrodes; interfering signals reduction; nerve cuff recordings; short circuited floating electrodes; traditional tripolar configuration; Biomedical measurements; Circuits; Electrodes; Electromyography; In vitro; Performance evaluation; Samarium; Signal to noise ratio; Silicon compounds; Wire;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    [Engineering in Medicine and Biology, 1999. 21st Annual Conference and the 1999 Annual Fall Meetring of the Biomedical Engineering Society] BMES/EMBS Conference, 1999. Proceedings of the First Joint
  • Conference_Location
    Atlanta, GA
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-5674-8
  • Type

    conf

  • DOI
    10.1109/IEMBS.1999.802555
  • Filename
    802555