• DocumentCode
    3474050
  • Title

    Feasibility study for SPECT mammography based on compact imagers rotating around breast vertical axis

  • Author

    Pani, R. ; Soluri, A. ; Scafè, R. ; Pellegrini, R. ; Vincentis, G. ; Cinti, M.N. ; Betti, M. ; Inches, R. ; Garibaldi, G. ; Cusanno, F. ; Gambaccini, M. ; Fantini, A. ; Taibi, A. ; Olivo, A. ; Pani, S. ; Rigon, L. ; Bollini, D. ; Lanconelli, N. ; Guerra,

  • Author_Institution
    Dept. of Expt. Med., Rome Univ., Italy
  • Volume
    3
  • fYear
    2000
  • fDate
    2000
  • Abstract
    The detection limit of invasive carcinoma by standard prone scintimammography appears to be >=1 cm diameter. Since it is desirable to detect lesions at very earliest stages of growth (0.8 cm or less), the development of prototype scintimammographic systems with improved imaging performances is a primary goal. Here, the authors propose a dedicated high resolution breast imaging scanner for SPECT around the Vertical Axis of Rotation (VAOR) with the breast in prone position. The rotating detector module consists of a compact position sensitive photomultiplier tube (PS-PMTHamamatsu R7600-00-C8 coupled to a CsI(Tl) scintillating array. The compactness is the peculiarity of this detector module to allow the lodging in the breast interspace to work close to the chest wall and to minimize the tumor to collimator front distance. A preliminary study was performed by a breast phantom and 5" small FOV gamma camera with an actual rotation radius of 6.5 cm. Images were reconstructured using a filtered back projection implementation. Tumor SNR from planar images shown very low values: on the contrary reconstructed images strongly enhanced the contrast of a tumor 1 cm sized. A further enhancement was obtained by the compact detector module that was able to achieve higher SNR values for tumor less 1 cm sized)
  • Keywords
    biomedical equipment; cancer; mammography; photomultipliers; single photon emission computed tomography; 0.8 cm; 1 cm; 6.5 cm; CsI(Tl) scintillating array; CsI:Tl; SPECT mammography; breast interspace; breast vertical axis; compact imagers; invasive carcinoma detection limit; medical diagnostic imaging; medical instrumentation; nuclear medicine; position sensitive photomultiplier tube; rotating detector module; Breast neoplasms; High-resolution imaging; Image reconstruction; Image resolution; Lesions; Mammography; Photomultipliers; Position sensitive particle detectors; Prototypes; Sensor arrays;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2000 IEEE
  • Conference_Location
    Lyon
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-6503-8
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2000.949342
  • Filename
    949342