Abstract :
The paper describes a new power sensors´ calibration method suitable for the twin type coaxial microcalorimeter. This method allows an intrinsic compensation of the lines losses, thus producing direct results, as in an ideal unitary-gain microcalorimeter. Furthermore, it does not require the microcalorimeter calibration step with the short circuit and an equation similar to that for the ideal system is applied for the measurand computation.
Keywords :
calibration; calorimeters; coaxial cables; microwave measurement; power measurement; sensors; switching; thermoelectric power; coaxial microcalorimeter; direct effective efficiency measurement; line losses; microcalorimeter calibration; power sensor calibration method; short circuit; switching method; thermoelectric power sensor; true-twin microcalorimeter; unitary-gain microcalorimeter; Calibration; Coaxial components; Dielectric measurements; Electromagnetic measurements; Equations; Hardware; Power measurement; Sensor phenomena and characterization; Thermal sensors; Thermoelectricity;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements (CPEM), 2010 Conference on
Conference_Location :
Daejeon
Print_ISBN :
978-1-4244-6795-2
Type :
conf
DOI :
10.1109/CPEM.2010.5544182
Filename :
5544182
Link To Document :
بازگشت