DocumentCode :
3499131
Title :
Influence of visualization parameters on linear size determination in CT-image
Author :
Tabakov, Slavik D.
Author_Institution :
Dept. of Roentgenol., Med. Acad.-Plovdiv, Bulgaria
fYear :
1988
fDate :
4-7 Nov. 1988
Abstract :
The visualization parameters (window, center, and filters) have great influence on the determination of linear dimensions of objects in computed tomography (CT). This is due to the spread of image contours (the density interface), which is from 1- to 6-mm wide. This is a significant source of errors in linear measurements, especially in small objects with high density gradient. A way of solving this problem by measuring from the denso-profile is proposed.<>
Keywords :
computerised tomography; length measurement; computed tomography; density interface; denso-profile; image contours; linear measurements; linear size determination; measurement errors; object dimensions; visualization parameters;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 1988. Proceedings of the Annual International Conference of the IEEE
Conference_Location :
New Orleans, LA, USA
Print_ISBN :
0-7803-0785-2
Type :
conf
DOI :
10.1109/IEMBS.1988.95128
Filename :
95128
Link To Document :
بازگشت