• DocumentCode
    3529944
  • Title

    Estimation of energy range measurements with newly developed Si/CdTe Compton camera for nuclear medicine imaging

  • Author

    Yamaguchi, Mitsutaka ; Kawachi, Naoki ; Kamiya, Tomihiro ; Suzui, Nobuo ; Fujimaki, Shu ; Odaka, Hirokazu ; Ishikawa, Shinnosuke ; Kokubun, Motohide ; Watanabe, Shin ; Takahashi, Tadayuki ; Shimada, Hirofumi ; Arakawa, Kazuo ; Suzuki, Yoshiyuki ; Torikai,

  • Author_Institution
    Takasaki Adv. Radiat. Res. Inst., Japan Atomic Energy Agency, Takasaki, Japan
  • fYear
    2010
  • fDate
    Oct. 30 2010-Nov. 6 2010
  • Firstpage
    1491
  • Lastpage
    1493
  • Abstract
    In this paper, we report the estimations of the lower detection limits of gamma ray energy for a newly developed Compton camera for applications to nuclear medicine imaging. We performed simulation studies to investigate the relationship of the distance between the scatterer and absorber detectors with the lower detection limit. In addition, imaging tests were performed by using multiple sources.
  • Keywords
    Compton effect; II-VI semiconductors; cadmium compounds; elemental semiconductors; radioisotope imaging; silicon; silicon radiation detectors; wide band gap semiconductors; Compton camera; Si-CdTe; absorber detectors; energy range measurements; gamma ray energy; lower detection limits; nuclear medicine imaging; scatterer detectors; Biomedical imaging; Cameras; Detectors; Gamma rays; Scattering; Silicon;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
  • Conference_Location
    Knoxville, TN
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-9106-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2010.5874024
  • Filename
    5874024