Title :
Maximum-likelihood calibration of an x-ray computed tomography system
Author :
Moore, Jared W. ; Van Holen, Roel ; Barrett, Harrison H. ; Furenlid, Lars R.
Author_Institution :
Coll. of Opt. Sci., Ghent Univ., Ghent, Belgium
fDate :
Oct. 30 2010-Nov. 6 2010
Abstract :
We present a maximum-likelihood (ML) method for calibrating the geometrical parameters of an x-ray computed tomography (CT) system. This method makes use of the full image data and not a reduced set of data. This algorithm is particularly useful for CT systems that change their geometry during the CT acquisition, such as an adaptive CT scan. Our ML search method uses a contracting-grid algorithm that does not require initial starting values to perform its estimate, thus avoiding problems associated with choosing initialization values.
Keywords :
calibration; computerised tomography; grid computing; maximum likelihood estimation; medical diagnostic computing; ML search method; X-ray computed tomography system; adaptive CT scan; contracting-grid algorithm; maximum-likelihood calibration; Calibration; Computed tomography; Maximum likelihood estimation; Mice; Trajectory; X-ray imaging;