DocumentCode :
3581576
Title :
Doppler radar noncontact imaging of human cardiac motion
Author :
Li, Changzhi ; Zhu, Anjie ; Qiao, Shan ; Sun, Yongzhi ; Ran, Lixin
Author_Institution :
Department of Electrical and Computer Engineering, Texas Tech University, Lubbock, 79409 USA
fYear :
2014
Firstpage :
941
Lastpage :
943
Abstract :
This paper reports recent investigation on the possibility of human cardiac motion imaging using a continuous-wave microwave Doppler radar sensor. Compared with existing works of vital sign radar sensors that mainly focused on the heartbeat rate, the challenge is detect the actual pattern of cardiac motion. To achieve this, a DC-coupled direct conversion architecture and a digital-IF architecture can be used because both of them can avoid frequency-dependent signal distortion. A digital-IF Doppler radar sensor employing an extended differentiate and cross-multiply (DACM) algorithm was adopted to perform a series of experiments that investigate the effectiveness of noncontact cardiac imaging from different orientations of a subject. Experimental results demonstrated that in addition to detecting heartbeat rate, Doppler radar could also image one-dimensional, time domain cardiac motion that fits well with the known physiological description.
Keywords :
Baseband; Demodulation; Doppler radar; Imaging; Radar imaging; Tuning; DC-coupled; Doppler radar sensor; cardiac motion imaging; digital-IF; extended DACM;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Microwave Conference (APMC), 2014 Asia-Pacific
Type :
conf
Filename :
7067585
Link To Document :
بازگشت