  • Abstract
    The knowledge of the water equivalent depth of an ionization chamber in a proton beam is important when reference dosimetry is performed in a region where depth dose gradients exist, such as in the entrance region of a non-modulated beam, or when measured depth dose curves have to be corrected. In this work, an analytical model is presented to calculate the water equivalent depth of spherical and cylindrical ionization chambers in a clinical proton beam, accounting for the ionization chamber´s cavity, wall, waterproofing sleeve and central electrode. For the theoretical situation of an air cavity without wall in water, this water equivalent depth equals the depth of the effective point of measurement. The model, which is a refinement of a previous model in which only the cavity and wall were taken into account, can be applied in a region where the depth dose varies approximately linear with depth. An expression to combine the model for a spherical and a cylindrical ionization chamber is given as well. This way, values for thimble chambers are obtained. The model is verified by Monte Carlo calculations using PTRAN. A further refinement allows calculation of the effective measurement depth in a region with non-linear gradients such as in the neighborhood of the Bragg peak. It is shown that this model can be used to convert a depth dose distribution measured with a thimble ionization chamber to the true depth dose distribution
  • Keywords
    Monte Carlo methods; biomedical equipment; dosimetry; ionisation chambers; radiation therapy; water; Bragg peak; Monte Carlo calculations; PTRAN; air cavity; analytical expressions; cavity; central electrode; clinical proton beam; clinical proton beam dosimetry; cylindrical ionization chamber; depth dose curves; depth dose gradients; effective water depth; entrance region; ionization chamber; nonlinear gradients; nonmodulated beam; spherical ionization chambers; thimble chambers; wall; water equivalent depth; waterproofing sleeve; Analytical models; Calibration; Dosimetry; Electrodes; Ionization chambers; Monte Carlo methods; Particle beam measurements; Particle beams; Performance analysis; Performance evaluation;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2000. Proceedings of the 22nd Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-6465-1
  • Type

    conf

  • DOI
    10.1109/IEMBS.2000.897854
  • Filename
    897854