DocumentCode
3593809
Title
An Electron-Tracking Compton imaging camera based on a gaseous TPC and a scintillation camera
Author
Ueno, K. ; Hattori, K. ; Kabuki, S. ; Kubo, H. ; Kurosawa, S. ; Miuchi, K. ; Nagayoshi, T. ; Nishimura, H. ; Okada, Y. ; Orito, R. ; Takada, A. ; Tanimori, T. ; Tsuchiya, K.
Author_Institution
Kyoto Univ., Kyoto
Volume
3
fYear
2007
Firstpage
1960
Lastpage
1964
Abstract
We are developing an electron-tracking Compton imaging camera (ETCC) based on a gaseous time projection chamber (TPC) and a scintillation camera. The ETCC detects the energy and the direction of the incident gamma ray using the information of the recoil electron and the scattered gamma ray. We have developed the ETCC with a detection volume of 23 times 28 times 15 cm3 which consists of the 23 times 28 times 15 cm3 gaseous TPC and 30 times 30 cm2 GSO(Ce) scintillation camera. We obtained the gamma-ray image and investigated the performances of the ETCC. The angular resolution measure (ARM) and the scatter plane deviation (SPD) are 6.1 degree and 64.5 degree (HWHM) at 662 keV, respectively, and the energy resolution is 18.0% (FWHM) at 662 keV.
Keywords
Compton effect; gamma-ray astronomy; gamma-ray detection; scintillation counters; time projection chambers; ARM; Compton telescope; ETCC; GSO(Ce) scintillation camera; SPD; TPC; angular resolution measure; electron-tracking Compton imaging camera; gamma-ray astronomy; gamma-ray image; gaseous time projection chamber; scatter plane deviation; Cameras; Electromagnetic scattering; Electrons; Energy resolution; Gamma ray detection; Gamma ray detectors; Particle scattering; Prototypes; Solid scintillation detectors; X-ray scattering; Compton Camera; TPC; gaseous detector; scintillation camera;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2007. NSS '07. IEEE