DocumentCode :
3629290
Title :
Modified time-of-flight based calibration approach for ultrasonic computed tomography
Author :
Adam Filipik;Jiri Jan;Igor Peterlik;Dusan Hemzal;Nicole Ruiter;Radovan Jirik
Author_Institution :
Dept. of Biomedical Eng., FEEC, Brno University of Technology, Czech Republic
fYear :
2008
Firstpage :
2181
Lastpage :
2184
Abstract :
In the previous paper [11], a method for geometrical and transducer-time-delay auto-calibration of an ultrasonic computed tomography (USCT) system has been described, aiming at calibration of individual ultrasonic (US) transducer positions. The present contribution describes a novel modification of the method utilizing the particular USCT system concept: the exactly known spatial relations among transducers grouped in each of the transducer array systems (TASes). The algorithms used for the calibration remain based on the principles similar to the global positioning system (GPS) navigation, however, the positions and orientations of complete TASes are calibrated, rather than individual positions of transducers. This way, the number of unknowns is substantially reduced while the number of available equations remains unchanged. Consequently, a solution substantially more robust with respect to measurement noise can be obtained based on this highly overdetermined equation system. The method is capable of calibrating the individual positions of all ultrasonic transducers via their positions in TASes as well as their individual time delays at once during sc. empty measurement, without a need for any particular arrangements, e.g. calibration phantoms.
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE
ISSN :
1094-687X
Print_ISBN :
978-1-4244-1814-5
Electronic_ISBN :
1558-4615
Type :
conf
DOI :
10.1109/IEMBS.2008.4649627
Filename :
4649627