DocumentCode :
3664651
Title :
A study of drift characteristic in arrayed flexible chlorine ion sensors with differential reference electrodes
Author :
Tong-Yu Wu; Shi-Chang Tseng; Jung-Chuan Chou; Yi-Hung Liao; Ruei-Ting Chen; Min-Siang Huang
Author_Institution :
Grad. Sch. of Mech. Eng., Nat. Yunlin Univ. of Sci. &
fYear :
2015
fDate :
6/1/2015 12:00:00 AM
Firstpage :
67
Lastpage :
70
Abstract :
The drift rate is change of voltage by time dependence, that is isothermal and fixed composition in the solution. In this study, we prepared arrayed flexible chlorine ion sensors which are fabricated by radio frequency sputtering technology and screen-printed technology to measure the drift in the different NaCl solutions, we obtained output voltage in variation concentrations from 10-5M to 1 M NaCl solutions over 12 hours, the drift rates of the chloride ion sensing devices are 1.122 mV/hr in 1 M NaCl solution, 1.192 mV/hr in 10-1 M NaCl solution, 2.869 mV/hr in 10-2 M NaCl solution, 1.529 mV/hr in 10-3M NaCl solution, 0.988 mV/hr in 10-4 M NaCl solution and 3.991 mV/hr in 10-5 M NaCl solution, respectively.
Keywords :
"Conferences","Electron devices","Solid state circuits"
Publisher :
ieee
Conference_Titel :
Electron Devices and Solid-State Circuits (EDSSC), 2015 IEEE International Conference on
Print_ISBN :
978-1-4799-8362-9
Type :
conf
DOI :
10.1109/EDSSC.2015.7285051
Filename :
7285051
Link To Document :
بازگشت