DocumentCode
3671958
Title
Fabrication of 60 channel microelectrode arrays for future use with cultured neuronal networks
Author
V. P. Gomes;A. D. Barros;J. B. D. Filho;S. Martinoia;A. Pasquarelli;J. W. Swart
Author_Institution
School of Electr. &
fYear
2015
Firstpage
1
Lastpage
4
Abstract
Stimulation and recording of nerve cells is a procedure used for several applications, such as clinical therapies, study of basic neuroscience, and neural prostheses. Microtechnology and advances in material science have allowed to produce more sophisticated devices and with more functions. This paper focuses on the fabrication of planar 60-channel MEAs. The electrical characterization of the noise level from the TiN electrodes showed good sensitivity to noise, compatible with commercial systems. These electrodes received an artificial electrocardiogram signal (ECG) from a function generator and registered the same input signal but with lower amplitude. Finally, both cyclic voltammetry curves of the produced MEA and the commercial MEA exhibited similar shape, but the amplitude of the first one was significantly higher than in MCS-MEA, with an order difference of magnitude.
Keywords
"Semiconductor device measurement","Monitoring","Biomedical monitoring","Electric variables measurement","Electrodes","Gain measurement"
Publisher
ieee
Conference_Titel
Microelectronics Technology and Devices (SBMicro), 2015 30th Symposium on
Type
conf
DOI
10.1109/SBMicro.2015.7298107
Filename
7298107
Link To Document