• DocumentCode
    3684615
  • Title

    Towards the development of a wearable Electrical Impedance Tomography system: A study about the suitability of a low power bioimpedance front-end

  • Author

    Matteo Menolotto;Stefano Rossi;Paolo Dario;Luigi Della Torre

  • Author_Institution
    Advanced System, Technology department, STMicroelectronics, Agrate Brianza, Italy
  • fYear
    2015
  • Firstpage
    3133
  • Lastpage
    3136
  • Abstract
    Wearable systems for remote monitoring of physiological parameter are ready to evolve towards wearable imaging systems. The Electrical Impedance Tomography (EIT) allows the non-invasive investigation of the internal body structure. The characteristics of this low-resolution and lowcost technique match perfectly with the concept of a wearable imaging device. On the other hand low power consumption, which is a mandatory requirement for wearable systems, is not usually discussed for standard EIT applications. In this work a previously developed low power architecture for a wearable bioimpedance sensor is applied to EIT acquisition and reconstruction, to evaluate the impact on the image of the limited signal to noise ratio (SNR), caused by low power design. Some anatomical models of the chest, with increasing geometric complexity, were developed, in order to evaluate and calibrate, through simulations, the parameters of the reconstruction algorithms provided by Electrical Impedance Diffuse Optical Reconstruction Software (EIDORS) project. The simulation results were compared with experimental measurements taken with our bioimpedance device on a phantom reproducing chest tissues properties. The comparison was both qualitative and quantitative through the application of suitable figures of merit; in this way the impact of the noise of the low power front-end on the image quality was assessed. The comparison between simulation and measurement results demonstrated that, despite the limited SNR, the device is accurate enough to be used for the development of an EIT based imaging wearable system.
  • Keywords
    "Tomography","Biomedical monitoring","Image reconstruction","Impedance","Electrodes","Current measurement","Phantoms"
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Electronic_ISBN
    1558-4615
  • Type

    conf

  • DOI
    10.1109/EMBC.2015.7319056
  • Filename
    7319056