DocumentCode :
3684783
Title :
Estimation of heart rate and heart rate variability from pulse oximeter recordings using localized model fitting
Author :
Federico Wadehn;David Carnal;Hans-Andrea Loeliger
Author_Institution :
Dept. of Information Technology and Electrical Engineering ETH Zurich, Switzerland
fYear :
2015
Firstpage :
3815
Lastpage :
3818
Abstract :
Heart rate variability is one of the key parameters for assessing the health status of a subject´s cardiovascular system. This paper presents a local model fitting algorithm used for finding single heart beats in photoplethysmogram recordings. The local fit of exponentially decaying cosines of frequencies within the physiological range is used to detect the presence of a heart beat. Using 42 subjects from the CapnoBase database, the average heart rate error was 0.16 BPM and the standard deviation of the absolute estimation error was 0.24 BPM.
Keywords :
"Electrocardiography","Heart beat","Heart rate variability","Oscillators","Estimation","Shape"
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN :
1094-687X
Electronic_ISBN :
1558-4615
Type :
conf
DOI :
10.1109/EMBC.2015.7319225
Filename :
7319225