DocumentCode
3685686
Title
Effects of stray capacitance to ground in three electrode monopolar needle bioimpedance measurements
Author
H. Kalvøy;C. Aliau-Bonet;R. Pallas-Areny;Ø.G. Martinsen
Author_Institution
Clinical and Biomedical engineering department at Rikshospitalet, Oslo University Hospital, Norway
fYear
2015
Firstpage
7542
Lastpage
7545
Abstract
Positive phase angle is documented and analyzed in a three electrode monopolar needle measurement. Inductance equivalent behavior of the stray capacitance to ground is described as error source in a non-inductive sample measurement.
Keywords
"Electrodes","Impedance","Impedance measurement","Capacitance","Needles","Frequency measurement","Biomedical measurement"
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN
1094-687X
Electronic_ISBN
1558-4615
Type
conf
DOI
10.1109/EMBC.2015.7320137
Filename
7320137
Link To Document