• DocumentCode
    3685686
  • Title

    Effects of stray capacitance to ground in three electrode monopolar needle bioimpedance measurements

  • Author

    H. Kalvøy;C. Aliau-Bonet;R. Pallas-Areny;Ø.G. Martinsen

  • Author_Institution
    Clinical and Biomedical engineering department at Rikshospitalet, Oslo University Hospital, Norway
  • fYear
    2015
  • Firstpage
    7542
  • Lastpage
    7545
  • Abstract
    Positive phase angle is documented and analyzed in a three electrode monopolar needle measurement. Inductance equivalent behavior of the stray capacitance to ground is described as error source in a non-inductive sample measurement.
  • Keywords
    "Electrodes","Impedance","Impedance measurement","Capacitance","Needles","Frequency measurement","Biomedical measurement"
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Electronic_ISBN
    1558-4615
  • Type

    conf

  • DOI
    10.1109/EMBC.2015.7320137
  • Filename
    7320137