Remote pulmonary function testing using a depth sensor
Author
Vahid Soleimani;Majid Mirmehdi;Dima Damen;Sion Hannuna;Massimo Camplani;Jason Viner;James Dodd
Author_Institution
Visual Information Laboratory, Faculty of Engineering, University of Bristol, BS8 1UB, UK
fYear
2015
Firstpage
1
Lastpage
4
Abstract
We propose a remote non-invasive approach to Pulmonary Function Testing using a time-of-flight depth sensor (Microsoft Kinect V2), and correlate our results to clinical-standard spirometry results. Given point clouds, we approximate 3D models of the subject´s chest, estimate the volume throughout a sequence and construct volume-time and flow-time curves for two prevalent spirometry tests: Forced Vital Capacity and Slow Vital Capacity. From these, we compute clinical measures, such as FVC, FEV1, VC and IC. We correlate automatically extracted measures with clinical spirometry tests on 40 patients in an outpatient hospital setting. These demonstrate high within-test correlations.
Keywords
"Volume measurement","Static VAr compensators","Lungs","Three-dimensional displays","TV","Testing","Cameras"
Publisher
ieee
Conference_Titel
Biomedical Circuits and Systems Conference (BioCAS), 2015 IEEE
Type
conf
DOI
10.1109/BioCAS.2015.7348445