DocumentCode
3748987
Title
ECG baseline wander removal with recovery of the isoelectric level
Author
Antonio Fasano;Valeria Villani
Author_Institution
Universit? Campus Bio-Medico di Roma, Rome, Italy
fYear
2015
Firstpage
577
Lastpage
580
Abstract
Baseline wander removal is an unavoidable step in ECG signal processing. The in-band nature of this noise makes its removal difficult without affecting the ECG. Many approaches have been proposed in the literature. Among them, cubic spline interpolation is the only one able to recover the isoelectric level in the detrended signal. However, it exhibits poor detrending performance. In this paper we extend our recent approach based on the notion of quadratic variation reduction, to address the problem of recovery of the isoelectric level. This is achieved by constraining the amplitude of few fiducial points to lie on isoelectric segments. Conversely to cubic spline interpolation, which requires a fiducial point for each beat, the proposed approach requires very few points: as few as one single point is sufficient. Simulation results show that the proposed approach largely outperforms cubic spline interpolation, being very effective in removing baseline wander and recovering the isoelectric level, while preserving ECG morphology.
Keywords
"Electrocardiography","Distribution functions","Splines (mathematics)","Simulation","Signal processing","Electrodes","Linear programming"
Publisher
ieee
Conference_Titel
Computing in Cardiology Conference (CinC), 2015
ISSN
2325-8861
Print_ISBN
978-1-5090-0685-4
Electronic_ISBN
2325-887X
Type
conf
DOI
10.1109/CIC.2015.7410976