DocumentCode
3759569
Title
The long object imaging problem in volumetric CT using iterative reconstruction algorithm, revisited
Author
Daxin Shi
Author_Institution
Toshiba Medical Research Institute of USA, Inc., 706 N. Deerpath Dr., Vernon Hills, IL, 60061, USA
fYear
2014
Firstpage
1
Lastpage
3
Abstract
The long object imaging problem in volumetric CT using iterative reconstruction algorithm is a challenge due to the intrinsic longitudinal (z-axis) truncation in the circular scanning trajectory. We revisited the problem in this work and proposed a simplified version of our original approach. We simply removed the projection data expansion and empirical weighting scheme in our original method while reconstructed only the image in an extended region. It seems that error will be introduced in the reconstructed image volume from those rays traveling both through the region of interest (ROI) and outside ROI. Hence error will propagate further into the image volume resulting the useful ROI diminishes for each iteration. However, our results showed that the image quality was well preserved by use of iterative reconstruction method. Artifacts due to the longitudinal truncation did not propagate as originally expected. We demonstrated our results of two sets of clinical data, head and abdomen, respectively, acquired from Aquilion ONE™ CT scanner (Toshiba American Medical Systems, Tustin, CA, USA) reconstructed by use of iterative reconstruction algorithm.
Publisher
ieee
Conference_Titel
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2014 IEEE
Type
conf
DOI
10.1109/NSSMIC.2014.7430802
Filename
7430802