DocumentCode :
3759659
Title :
Large view x-ray imaging for oversized sample with the grating interferometry
Author :
Xiaohua Zhu; Li Zhang; Xiaolei Jiang; Ran Zhang
Author_Institution :
Department of Engineering Physics, Tsinghua University, Beijing 100084, China
fYear :
2014
Firstpage :
1
Lastpage :
4
Abstract :
X-ray grating interferometry, as a powerful nondestructive imaging tool in fields like biology and medical tests, works well with conventional X-ray tubes, and provides three kinds of information, the attenuation image, the phase-contrast image and the dark field image, with higher contrast. However, the limited field of view of the sample due to the limited size of gratings and detector hinders the further application of the technique. In this paper, we report a solution of large view imaging for objects oversized comparing to the gratings utilizing 2D translation of the sample, together with image-based methods as image alignment and weighted blending to overcome the difficulty from the inaccuracy of the translational devices and exposure responses, to obtain the full view of the sample object in one image, providing a possible future application of large size object scan with accessible device accuracy. Experimental results verify the feasibility of the method.
Keywords :
"Gratings","X-ray imaging","Imaging","Interferometry","Detectors","Electron tubes","Attenuation"
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2014 IEEE
Type :
conf
DOI :
10.1109/NSSMIC.2014.7430891
Filename :
7430891
Link To Document :
بازگشت