3″ × 3″ LaBr3:Ce position sensitivity with multi-anode PMT readout
Author
Agnese Giaz;Nives Blasi;Franco Camera;Ciro Boiano;Sergio Brambilla;Benedicte Million;Stefano Riboldi
Author_Institution
Istituto Nazionale di Fisica Nucleare INFN, Sezione di Milano, Via Celoria 16, 20133 (Mi), Italy
fYear
2014
Firstpage
1
Lastpage
5
Abstract
The response of a cylindrical continuous 3" × 3" LaBr3:Ce crystal coupled to four Position Sensitive photomultipliers (PSPMT) was studied using a 1 mm collimated beam of 662 keV gamma rays. The LaBr3:Ce crystal is sealed in a capsule and has diffusive/reflective surfaces, the energy resolution is 21 keV (FWHM) at 662 keV, measured using a R6233-100sel PMT. The PSPMTs are HAMAMATSU H8500 and have 64 segments each. Two PSPMTS have 10 dynodes and a 52 mm flat square standard Bialkali photocathode while the other two have 8 dinodes and a SBA photocathode. We performed a scan of the crystal moving the collimated source. The 256 segments of the four PSPMTs were grouped into 4 or 16 elements. The Centroid position, FWHM and area of the full energy peak were extracted from the measured spectra. The data show a regular behavior which depends on the position of the incident gamma beam. The position of the source was also calculated event by event and the positional resolution was measured.
Keywords
"Crystals","Energy resolution","Energy measurement","Gamma-rays","Cameras","Position measurement","Detectors"