• DocumentCode
    3767372
  • Title

    A design of 2nd order DT sigma-delta modulator for medical implants

  • Author

    MA Raheem;K. Manjunatha Chari;Mohammed Arifuddin Sohel;M A Mushahhid Majeed

  • Author_Institution
    ECED, Muffakham Jah College of Engineering and Technology, Hyderabad, India
  • fYear
    2015
  • Firstpage
    6
  • Lastpage
    10
  • Abstract
    This paper proposes the design of a second order sigma delta modulator (SDM) to be used in medical implants for conversion of low frequency human body analog signals into digital format. The lifetime of a medical implant is governed by the power consumed by various blocks in it, of which the block consuming maximum power is the Analog to Digital Converter (ADC). This paper aims at designing a power optimized ADC by making use of CMOS operational trans-conductance amplifier(OTA) whose transistors are biased in sub threshold conduction region by making use of body biasing technique. The modulator thus implemented consumes a power of 4.6 μW from a 0.9V power supply. The modulator achieves an SNR of 64dB for a bandwidth of 500 Hz.
  • Keywords
    "Implants","CMOS integrated circuits","CMOS technology","Modulation","Switches","Gain","Signal to noise ratio"
  • Publisher
    ieee
  • Conference_Titel
    Microelectronics and Electronics (PrimeAsia), 2015 IEEE Asia Pacific Conference on Postgraduate Research in
  • Electronic_ISBN
    2159-2160
  • Type

    conf

  • DOI
    10.1109/PrimeAsia.2015.7450460
  • Filename
    7450460