• DocumentCode
    377649
  • Title

    Diagnostics plate for SNS linac commissioning

  • Author

    Plum, M.A. ; Garnett, R.W. ; Meyer, R., Sr. ; Shafer, R.E.

  • Author_Institution
    Los Alamos Nat. Lab., NM, USA
  • Volume
    3
  • fYear
    2001
  • fDate
    2001
  • Firstpage
    2374
  • Abstract
    A diagnostics plate is being assembled as part of the linac commissioning plan for the Oak Ridge National Laboratory Spallation. Neutron Source (SNS). The D-Plate will be positioned directly after the first drift tube linac; (DTL) tank, where the H- beam energy will be 7.5 MeV. A full suite of beam diagnostic instrumentation is planned, including profile, current, phase, energy, and emittance measurements. The D-Plate is designed to operate up to the full duty factor of the linac, or up to 16-kW beam power. Until the spallation neutron target is completed several years later, this will be the last time the facility can be operated at full current and duty cycle
  • Keywords
    ion accelerators; linear accelerators; particle beam diagnostics; 16 kW; 7.5 MeV; D-Plate; H; Spallation. Neutron Source; beam diagnostics; current measurement; diagnostics plate; emittance measurement; energy measurement; linac; phase measurement; Apertures; Current measurement; Energy measurement; Instruments; Linear particle accelerator; Magnets; Monitoring; Position measurement; Power measurement; Wire;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Particle Accelerator Conference, 2001. PAC 2001. Proceedings of the 2001
  • Conference_Location
    Chicago, IL
  • Print_ISBN
    0-7803-7191-7
  • Type

    conf

  • DOI
    10.1109/PAC.2001.987386
  • Filename
    987386